Pulmonary vein isolation by duty-cycled bipolar and unipolar radiofrequency energy with a multielectrode ablation
Lucas V A Boersma1, Maurits C E F Wijffels, Hakan Oral
1Cardiology Department, St. Antonius Hospital, Nieuwegein, The Netherlands. l.boersma@antonius.net
Background:
Pulmonary vein (PV) isolation for ablation of atrial fibrillation (AF) remains a complex and lengthy procedure.
Objective:
The purpose of this study was to evaluate the feasibility and safety of a novel multielectrode catheter that delivers duty-cycled bipolar and unipolar radiofrequency (RF) energy.
Methods:
Patients eligible for catheter ablation of paroxysmal AF after screening with magnetic resonance imaging and transesophageal echocardiography were included in the study. A decapolar (3-mm electrode, 3-mm spacing, 25-mm diameter), circular, over-the-wire mapping and ablation catheter was deployed in the antrum of each PV. Ablation was performed with 60-second, 60 degrees C applications of duty-cycled bipolar/unipolar RF in a 4:1 ratio simultaneously at all selected electrode pairs until local activity was no longer observed. At 6 months, 7-day Holter monitoring was performed to determine freedom from AF without use of antiarrhythmic drugs.
Results:
In 98 patients (mean age 59 +/- 9 years), the PV ablation catheter was used for ablation of 369 veins (20 common left antra). All targeted veins (100%) were isolated as confirmed by the absence of potentials in the ostium either by PV ablation catheter or Lasso mapping. Mean number of RF applications was 27 +/- 7, total procedural time 84 +/- 29 minutes, and fluoroscopy time 18 +/- 8 minutes. Follow-up after 6 months without antiarrhythmic drugs showed freedom from AF in 83% of patients. No procedure-related complications were observed.
Conclusion:
PV isolation by duty-cycled bipolar/unipolar low-power RF energy through a circular, decapolar catheter can be achieved safely and efficiently, with good efficacy at 6 months.