Reliability of Artificial Intelligence-Based Cone Beam Computed Tomography Integration with Digital Dental Images
Published on: February 23, 2024
Image quality analysis to reduce dental artifacts in head and neck imaging with dual-source computed tomography
D Ketelsen1, M K Werner, C Thomas
1Abteilung für Diagnostische und Interventionelle Radiologie, Eberhard-Karls-Universität Tübingen. dominik.ketelsen@med.unituebingen.de
Summary
Dual-source computed tomography (DSCT) neck imaging with multiplanar reconstructions (MPRs) shows significant dental artifacts. An additional angulated spiral significantly reduces these artifacts, improving image quality for oropharyngeal structures.
Area of Science:
- Radiology
- Medical Imaging
- Otorhinolaryngology
Background:
- Dental implants frequently cause metallic artifacts in medical imaging.
- These artifacts can obscure important oropharyngeal structures in neck CT scans.
- Dual-source computed tomography (DSCT) is a common imaging modality for the neck.
Purpose of the Study:
- To compare the effectiveness of multiplanar reconstructions (MPRs) versus an angulated spiral technique in reducing dental artifacts in DSCT of the neck.
- To evaluate the impact of these techniques on the image quality of oropharyngeal structures.
Main Methods:
- Sixty-two patients undergoing DSCT neck imaging were analyzed.
- Image quality was assessed using both MPRs from axial datasets and an additional short spiral acquired parallel to the mouth floor.
- Anatomical structures were evaluated for artifact severity on a scale of 1 (least) to 4 (most).
Main Results:
- MPRs resulted in significant artifacts in 87.1% of anatomical structures (mean score 3.12 +/- 0.86).
- The angulated spiral technique showed negligible artifacts in leading oropharyngeal structures (mean score 1.28 +/- 0.46).
- The reduction in artifact severity with angulated slices was statistically significant (p < 0.01).
Conclusions:
- MPRs are insufficient for adequately reducing dental artifacts in DSCT of the neck.
- An additional short angulated spiral, parallel to the floor of the mouth, is recommended for routine use in patients with dental implants.
- Utilizing a flexible head holder is crucial due to the static gantry design of DSCT.
Related Concept Videos
Computed Tomography
Tomography refers to imaging by sections. Computed tomography (CT) is a non-invasive imaging technique that uses computers to analyze several cross-sectional X-rays to reveal minute details about structures in the body.
The technique was invented in the 1970s and is based on the principle that as X-rays pass through the body, they are absorbed or reflected at different levels. In the technique, a patient lies on a motorized platform while a computerized axial tomography (CAT) scanner rotates...
The technique was invented in the 1970s and is based on the principle that as X-rays pass through the body, they are absorbed or reflected at different levels. In the technique, a patient lies on a motorized platform while a computerized axial tomography (CAT) scanner rotates...
Imaging Studies III: Computed Tomography
DefinitionComputed Tomography (CT) of the genitourinary (GU) tract is a non-invasive imaging modality that utilizes X-rays and computer processing to generate detailed cross-sectional images of the urinary system, encompassing the kidneys, ureters, bladder, and adjacent structures such as the adrenal glands.PurposeCT scans of the GU tract serve several diagnostic and therapeutic purposes, including:Diagnosis of Urinary Tract Diseases: Detects kidney stones, tumors, cysts, and congenital...