Background:
Currently, one or two dosimeters are used to monitor radiation exposure in most cardiac laboratories. In addition, several different formulas are used to convert exposure data into an effective dose (ED).
Purpose:
To clarify the effect of monitoring methods and formula selection on the estimated ED for physicians during percutaneous coronary interventions (PCIs).
Material And Methods:
The ED of physicians during cardiac catheterization was determined using an optically stimulated luminescence dosimeter (Luxel badge). Two Luxel badges were worn: one beneath a personal lead apron (0.35-mm lead equivalent) at the chest and one outside of the apron at the neck.
Results:
The difference in the average ED of seven physicians was approximately fivefold (range 1.13-5.43 mSv/year) using the six different formulas in the clinical evaluation. The estimated physician ED differed markedly according to both the monitoring method and formula selected.
Conclusion:
ED estimation is dependent on both the monitoring method and the formula used. Therefore, it is important that comparisons among laboratories are based on the same monitoring method and same formula for calculating the ED.
Related Concept Videos
Biological Effects of Radiation
Radiological Investigation III: Pulmonary Angiogram and PET Scan
Pulmonary Angiogram
A Pulmonary Angiogram is an invasive procedure involving injecting a contrast medium through a catheter threaded into the pulmonary artery or the right side of the heart to visualize the pulmonary vasculature. Computed Tomography (CT) scans have mainly replaced this...
Cardiac Catheterization I: Pre-Procedure Overview
Imaging Studies for Cardiovascular System III: X-Ray
Definition and Purpose
An X-ray, or radiograph, is a non-invasive method that uses ionizing radiation to take images of internal structures. It is mainly used in cardiac imaging to examine the heart, lungs, and major blood vessels, aiming to identify abnormalities in the heart's size, shape, and position, such as heart failure, congenital defects, and vascular...
Imaging Studies for Cardiovascular System V: CT