Related Experiment Video
Updated: Jun 27, 2026

Computed Tomography-guided Time-domain Diffuse Fluorescence Tomography in Small Animals for Localization of Cancer Biomarkers
Published on: July 17, 2012
Investigation of detection limits for diffuse optical tomography systems: II. Analysis of slab and cup geometry for
Ronny Ziegler1, Bernhard Brendel, Herbert Rinneberg
1Philips Research Europe-Hamburg, Röntgenstr. 24, 22335 Hamburg, Germany.
Abstract:
Using a statistical (chi-square) test on simulated data and a realistic noise model derived from the system's hardware we study the performance of diffuse optical tomography systems for fluorescence imaging. We compare the predicted smallest size of detectable lesions at various positions in slab and cup geometry and model how detection sensitivity depends on breast compression and lesion fluorescence contrast. Our investigation shows that lesion detection is limited by relative noise in slab geometry and by absolute noise in cup geometry.