Related Experiment Video
Updated: Jun 26, 2026

Invasive Hemodynamic Assessment for the Right Ventricular System and Hypoxia-Induced Pulmonary Arterial Hypertension in Mice
Published on: October 24, 2019
Changes in right ventricular pressures between hemodialysis sessions recorded by an implantable hemodynamic monitor
Barbro Kjellström1, Frieder Braunschweig, Erland Löfberg
1Medtronic, Inc., Minneapolis, Minnesota, USA. barbro.kjellstrom@medtronic.com
Abstract:
Intermittent and chronic volume overload might contribute to the onset and progression of cardiovascular disease in patients who are undergoing maintenance hemodialysis (HD). Continuous monitoring of central hemodynamic variables may provide valuable information to improve volume control, particularly in patients with left ventricular dysfunction. Sixteen patients with end-stage renal disease who were undergoing long-term HD received an implantable hemodynamic monitor consisting of a subcutaneously implanted memory device and transvenous right ventricular (RV) lead with a pressure sensor. The implantable hemodynamic monitor continuously records heart rate, RV pressures, and estimated pulmonary arterial (PA) diastolic pressure, an estimate of left ventricular filling pressure. All patients underwent HD 3 times per week, and averages of rest hemodynamic values from the first, second, and third nights after HD during 12 weeks were analyzed. The third night always occurred after the weekend, when there was an extended interval between dialysis sessions. From the first night to the second night, RV systolic pressure increased by 10 +/- 8% (p <0.001), and estimated PA diastolic pressure increased by 16 +/- 14% (p <0.001). On the third night, RV systolic pressure increased by 14 +/- 12% (p <0.001), and estimated PA diastolic pressure increased by 23 +/- 18% (p <0.001) compared with the first night. In conclusion, the progressive pressure increments between dialysis sessions seen in this study suggest that the implantable hemodynamic monitor was a sensitive indicator for changes in volume load in patients who were undergoing HD treatment. The results also suggest that more frequent dialysis may avoid excessive pressure increase, but this needs to be investigated further in future studies.
Related Concept Videos
Hemodialysis II: Procedure and Complications
Hemodialysis III: Nursing Management
Hemodialysis I: Introduction
Cardiac Catheterization II: Right Heart Catheterization
Special considerations while measuring blood pressure
Monitoring Both Arms:
Monitoring BP in both arms during the initial assessment is advisable, as the systolic value may differ by five to ten mm Hg between arms. For subsequent BP assessments, use the arm with the higher reading.