Creatine-kinase MB mass: age and sex-associated reference limits in two different platforms that use the same method
M Zaninotto1, M M Mion, E Novello
1Department of Laboratory Medicine, University-Hospital, Padova, Italy. martina.zaninotto@sanita.padova.it
Clinica Chimica Acta; International Journal of Clinical Chemistry
|December 23, 2008
Summary
This study found that CK-MB mass levels are higher in men than women, a key finding for diagnosing myocardial necrosis when troponin assays are unavailable. The two analytical platforms showed comparable results for CK-MB mass determination.
Area of Science:
- Clinical Chemistry
- Biomarker Analysis
- Cardiovascular Diagnostics
Background:
- Troponins are preferred for myocardial necrosis detection per NACB guidelines.
- Creatine kinase-MB (CK-MB) mass remains a viable alternative biomarker when troponin assays are unavailable.
- Establishing reliable gender-specific reference ranges for CK-MB mass is crucial for accurate clinical interpretation.
Purpose of the Study:
- To compare the reliability of two analytical platforms in determining gender-specific 99th percentile values for CK-MB mass.
- To evaluate the impact of gender on CK-MB mass levels in healthy individuals.
- To assess age-related trends in CK-MB mass.
Main Methods:
- Serum samples from healthy individuals were analyzed in two laboratories using Beckman Coulter Access((R)) CKMB methods on UniCel DxI 800 and Access((R)) 2 analyzers.
- CK-MB mass was measured in micrograms per liter (microg/L).
- Plasma samples were also evaluated in one laboratory to compare with serum values.
Main Results:
- Both analytical platforms demonstrated comparable precision, with total imprecision (CV%) ranging from 5.02% to 9.05%.
- A statistically significant gender difference was observed, with higher CK-MB mass values in men (99th percentile: 7.64 microg/L) compared to women (5.19 microg/L).
- Lower median CK-MB mass values were found in younger subjects (18-28 years) compared to older age groups, a trend consistent in both serum and plasma samples.
Conclusions:
- The two analytical platforms provide comparable results for CK-MB mass measurement.
- Significantly higher CK-MB mass values in males compared to females are a critical consideration for patient classification in suspected myocardial necrosis.
- Current commercially available assays often lack this essential gender-specific information, highlighting a gap in diagnostic practice.