Criteria for opiate identification using liquid chromatography linked to tandem mass spectrometry: problems in
Elizabeth J Fox1, Shirley Twigger, Keith R Allen
1Department of Clinical Biochemistry, Leeds Teaching Hospitals NHS Trust, Britannia House, Morley, Leeds LS27 0DQ, UK. Elizabeth.Fox@leedsth.nhs.uk
Background:
Liquid chromatography linked to tandem mass spectrometry (LC/MS/MS) is being increasingly used for drug confirmation. At present, no official criteria exist for drug identification using this technique although the European Union (EU) criteria for compound identification have been adopted. These criteria are evaluated with respect to opiate confirmation by LC/MS/MS and problems highlighted.
Methods:
Urine samples screened positive for opiates by immunoassay were subjected to confirmation by LC/MS/MS using multiple reaction monitoring (MRM) and two separate buffer systems of pH 6.8 and 8.0, respectively. The EU criteria for compound identification were applied for confirmation of morphine, 6-monoacetylmorphine (6MAM), codeine and dihydrocodeine (DHC).
Results:
Using the pH 6.8 buffer, confirmation could be achieved for 84%, 94%, 96% and 95%, respectively, for samples demonstrating MRM chromatographic peaks at retention times for morphine, 6MAM, codeine and DHC. Failure to meet the EU criteria was mainly attributed to low signal-to-noise (S:N) ratios or excessively high drug concentrations. Isobaric interferences and poor chromatography were also contributing factors. The identification of morphine was considerably improved with chromatography at pH 8.0 owing to resolution of interferences. Oxycodone metabolites were a potential problem for the identification of DHC.
Conclusion:
Isobaric interferences can pose a problem with drug identification using LC/MS/MS. Optimizing chromatographic conditions is important to overcome these interferences. Consideration needs to be given to investigating drug metabolites as well as parent drugs in method development.
Related Concept Videos
High-Performance Liquid Chromatography: Types of Detectors
Mass Spectrometry: Complex Analysis
GC–MS is a powerful hyphenated method commonly used in forensics and environmental...
High-Performance Liquid Chromatography: Elution Process
Peptide Identification Using Tandem Mass Spectrometry
This technique helps gather information regarding the protein from which the peptide was obtained and to study the peptides’ amino acid sequence. Identifying peptides from a complex mixture is an important component of the growing field of...
High-Performance Liquid Chromatography: Introduction
In HPLC, two phases play a critical role in the separation process: