Related Experiment Videos
Internal supply of coenzyme to an amperometric glucose biosensor based on a chemically modified electrode
1Department of Analytical Chemistry, University of Lund, Sweden.
Abstract:
A biosensor for glucose using glucose dehydrogenase immobilized on a chemically modified graphite electrode was supplied with coenzyme, nicotinamide adenine dinucleotide (NAD+), through pores in the material. A graphite rod was hollowed out, leaving 0.3 mm at the end contacting the solution, filled with 10 mM NAD+ and pressurized. The response factor was 40% of that obtained when 2 mM NAD+ was mixed with the sample solution in a flow system. The coenzyme consumption was 11 microliters h-1 representing a 500-fold saving compared to supply through the bulk solution. The biosensor had a linear calibration curve from the detection limit, 1 microM, to 2 mM glucose and a repeatability of 0.3%. The graphite electrode was modified by adsorption of a bis-(benzophenoxazinyl)-terephthaloyl derivative in order to be able to oxidize NADH at 0 mV versus Ag/AgCl, 0.1 M KCl.