Imaging of Biological Tissues by Desorption Electrospray Ionization Mass Spectrometry
Published on: July 12, 2013
The influence of material and mesh characteristics on transmission mode desorption electrospray ionization
Joseph E Chipuk1, Jennifer S Brodbelt
1Department of Chemistry and Biochemistry, University of Texas at Austin, Austin, Texas 78712, USA.
Abstract:
Adaptation of desorption electrospray ionization to a transmission mode (TM-DESI) entails passing an electrospray plume through a sample that has been deposited onto a mesh substrate. A combination of mass spectrometry and fluorescence microscopy studies is used to illustrate the critical role material composition, mesh open space, and mesh fiber diameter play on the transmission, desorption, and ionization process. Substrates with open spaces less than 150 microm and accompanying minimal strand diameters produce less scattering of the plume and therefore favor transmission. Larger strand diameters typically encompass larger open spaces, but the increase in the surface area of the strand increases plume scattering as well as solvent and analyte spreading on the mesh. Polypropylene (PP), ethylene tetrafluoroethylene (ETFE), and polyetheretherketone (PEEK) materials afford much better desorption than similarly sized polyethylene terephthalate (PETE) or nylon-6,6 (PA66) substrates. Ultimately, the manner in which the electrospray plume interacts with the mesh as it is transmitted through the substrate is shown to be critical to performing and optimizing TM-DESI analyses. In addition, evidence is presented for analyte dependent variations in the desorption mechanisms of dry and solvated samples.
Related Concept Videos
Electrospray Ionization (ESI) Mass Spectrometry
ESI utilizes electrical energy to transfer ions from the liquid phase of the sample into the...
Mass Spectrum: Interpretation
Chemical Ionization (CI) Mass Spectrometry
Ion-Exchange Chromatography
Mass Spectrometers
Mass Spectrometry: Overview