Albumin-based or albumin-linked calibrators cause a positive bias in serum proteins assayed by the biuret method
Vratislav Chromý1, Lenka Sváchová, Lukás Novosád
1Institute of Chemistry, Faculty of Science, Masaryk University, Brno, Czech Republic. vrch@chemi.muni.cz
Background:
Assay of total serum protein by the biuret method calibrated with albumin standards according to the reference method provides results with a positive bias approximately 3%-5% exceeding the total error of 3.4% allowable for total protein in serum analysis made by analysers using two-part reagents and short-term procedures.
Methods:
We used two types of two-part biuret reagents utilised in a short-term measurement in analysers with albumin or serum calibrators, in which protein was attested by the Kjeldahl method.
Results:
Tests with potentially interfering substances proved that serum blanking used in a short-term biuret procedure is not capable of sufficiently eliminating effects of serum interferents. A short-term blanking is evidently capable of suppressing only an absorbance caused by serum-present coloured and turbid interferents, but its capacity to transform them (oxidise, hydrolyse, saponify, etc.) to some other not-interfering substances is very low compared with a long-term blanking. Lipids and bilirubin are responsible for significant positive bias of total protein in normal serum samples (approximately 3%) and even a greater positive offset in lipaemic and icteric sera (approximately 5%). We verified that interference tests based on a normal serum spiked with endogenous lipids and bilirubin give quite false and misleading results in the biuret reaction. A pure albumin, not depending on its bovine/human origin, gives absorbance responding only to its copper complexes with protein with a biuret regent, while its absorbance with a serum also includes the absorbance of interferents present in serum. The simplest way to improve current short-term biuret procedures is the use of a human serum calibrator with total protein attested by the Kjeldahl method. A serum calibrator, behaving analogously to serum samples, compensates for a positive bias in most normal sera. Reagents with a greater concentration of active biuret components (copper and alkali, reference method included) seem to be unnecessarily aggressive to proteins and are responsible for a lower accuracy when used in short-term measurements.
Conclusions:
Standard Reference Material 927c based on pure bovine albumin is still recommended and used as the primary standard for assays of total protein by colourimetric methods. The albumin calibrator is responsible for a positive bias of approximately 3%-5% in serum total protein assayed by the biuret reaction both in the reference and in current methods. Its substitution by a serum calibrator attested by the Kjeldahl method could solve this drawback. Clin Chem Lab Med 2009;47:91-101.
Related Concept Videos
Serum Studies: Renal Function Tests
Instrument Calibration
Analytical Balance Calibration
An analytical balance measures mass and requires regular calibration to...
Drug Distribution: Plasma Protein Binding
Enzyme-Linked Immunosorbent Assay
There are many different types of ELISAs, but they all involve an antibody molecule whose constant region binds an enzyme, leaving the variable region free to bind its specific antigen. Enzyme-substrate reaction allows the antigen to be visualized or quantified.
Western Blotting
The technique begins with separating proteins from the sample using sodium dodecyl sulfate-polyacrylamide gel electrophoresis (SDS-PAGE), followed by protein transfer, immunoblotting, and finally, protein detection.