Validation of the urine column measurement as an estimation of the intra-abdominal pressure
Herjan van der Steeg1, Jesse P van Akkeren, Saskia Houterman
1Department of Surgery, University Medical Centre St. Radboud, Nijmegen, The Netherlands. h.vandersteeg@chir.umcn.nl
Objective:
To evaluate the efficacy of the urine column (UC) measurement compared to the intra-vesicular pressure (IVP) measurement as an estimation of intra-abdominal pressure (IAP) in patients with IAP up to 30 mmHg.
Methods:
Fifteen patients undergoing a laparoscopic cholecystectomy were studied. All patients were catheterized. IVP measurements were performed using a pressure transducer connected to the culture aspiration port. UC measurements were done by holding up the tubing against a measuring rod. The symphysis pubis was used as the zero-reference. IAP was raised from 0 to 30 mmHg using increments of 5 mmHg, during which first the IVP and then UC measurement series were recorded end-expiratory. Fifty and 100 ml of saline were used as a priming volume.
Results:
The IVP and UC measurements showed a significant correlation with IAP. Comparing IVP and UC showed a correlation of 0.91 (p < 0.001) for 50 ml and 0.87 (p < 0.001) for 100 ml of saline as a priming volume. Using 50 ml of saline, UC was 0.68 mmHg higher than IVP (95% CI -7.21 to +5.85 mmHg). For 100 ml of saline, UC was 1.23 mmHg lower than IVP (95% CI -7.41 to +9.87 mmHg).
Conclusion:
UC measurement shows significant correlation to IVP measurement as an estimation of the IAP. Further study needs to be done to conclude whether UC measurement is a reliable clinical alternative to IVP measurement.
Related Concept Videos
Measurement of Fluid Pressure
A basic form of manometer is the piezometer, a vertical tube open at the top and filled with the same...
Urodynamic Studies: Uroflowmetry
Nursing Assessment of the Genitourinary System III: Percussion and Auscultation
Sites for measuring blood pressure
The Brachial Artery: Primary Site for Blood Pressure Measurement
Nursing Assessment of the Genitourinary System II: Inspection and Palpation
Assessment of blood pressure in brachial artery(two-step method)