An Experimental Protocol for Assessing the Performance of New Ultrasound Probes Based on CMUT Technology in Application to Brain Imaging
Published on: September 24, 2017
Forward-looking intracardiac ultrasound imaging using a 1-D CMUT array integrated with custom front-end electronics
Amin Nikoozadeh1, Ira O Wygant, Der-Song Lin
1Edward L. Ginzton Laboratory, Stanford University, Stanford, CA 94305, USA. aminn@stanford.edu
Summary
We developed a novel forward-looking intracardiac echocardiography (ICE) catheter using capacitive micromachined ultrasonic transducers (CMUTs) and custom integrated circuits. This technology enhances cardiac electrophysiology (EP) procedure guidance by improving visualization and reducing fluoroscopy use.
Area of Science:
- Biomedical Engineering
- Medical Imaging
- Ultrasound Technology
Background:
- Minimally invasive catheter-based electrophysiological (EP) interventions are standard for cardiac arrhythmias.
- Nonfluoroscopic intracardiac echocardiography (ICE) catheters improve EP guidance and reduce fluoroscopy.
- Current ICE catheters offer only side-looking imaging, limiting visualization.
Purpose of the Study:
- To develop a forward-looking ICE catheter for enhanced visualization during EP procedures.
- To integrate a 1-D CMUT array with custom transmit/receive circuitry for intracardiac imaging.
- To demonstrate the imaging capabilities of the developed forward-looking ICE catheter.
Main Methods:
- Fabrication of a 24-element, fine-pitch 1-D CMUT array (1.73 mm x 1.27 mm).
- Design of a custom integrated circuit (IC) with 24 transmit/receive channels (2.1 mm x 2.1 mm).
- Flip-chip bonding of the CMUT array to the IC, followed by PDMS coating and testing.
Main Results:
- Successful fabrication and integration of the CMUT array and custom IC.
- Achieved a pulse-echo center frequency of 9.2 MHz with 96% fractional bandwidth in water.
- Demonstrated imaging capability on phantoms and an ex vivo rabbit heart model.
Conclusions:
- The developed forward-looking ICE catheter shows promise for improved visualization in cardiac EP procedures.
- Integration of CMUTs and custom ICs enables high-performance intracardiac ultrasound imaging.
- This technology has the potential to reduce reliance on fluoroscopy and enhance procedural safety.