Cell-Free DNA Integrity Analysis in Urine Samples
Published on: January 5, 2017
Sample preparation module for bacterial lysis and isolation of DNA from human urine
M Dominika Kulinski1, Madhumita Mahalanabis2, Sara Gillers3
1Department of Mechanical Engineering, Boston University, 44 Cummington St., Boston, MA, USA.
Abstract:
Silica impregnated polymer monolithic columns may provide a simple method for lysing and extracting DNA from bacteria inside of microfluidic chips. Here we use Escherichia coli as a test organism for a point of care thermoplastic microfluidic module designed to take in a urine sample, mix it with lysis buffer, and perform a hybrid chemical/mechanical lysis and solid phase extraction of nucleic acids from the sample. To demonstrate proof-of-concept, we doped human hematuric urine samples with E. coli at concentrations ranging from 10(1)-10(5) colony-forming units/mL (CFU/mL) to simulate patient samples. We then performed on-chip lysis and DNA extraction. The bacterial DNA was amplified using real-time PCR demonstrating lysis and isolation down to 10(1) CFU/mL. Results were comparable to a commercial kit at higher concentrations and performed better at recovering DNA at lower concentrations.