Related Experiment Video
Updated: Jun 26, 2026

Thrombus Profiling Assay: A Microfluidics-Based Platform for Comprehensively Characterizing Biomechanical Thrombogenesis
Published on: January 9, 2026
Laboratory detection of the antiphospholipid syndrome via calibrated automated thrombography
Katrien Devreese1, Kathelijne Peerlinck, Jef Arnout
1Coagulation Laboratroy, Department of Clinical Chemistry, Microbiology and Immunology, Ghent University Hospital, Ghent, Belgium.
Abstract:
Lupus anticoagulants (LAC) consist of antiphospholipid antibodies, detected via their anticoagulant properties in vitro. Strong LAC relate to thromboembolic events, a hallmark of the antiphospholipid syndrome. We have analyzed whether detection of this syndrome would benefit from thrombin generation measurements. Therefore, calibrated automated thrombography was done in normal plasma (n = 30) and LAC patient plasma (n = 48 non-anticoagulated, n = 12 on oral anticoagulants), diluted 1:1 with a normal plasma pool. The anti-beta2-glycoprotein I monoclonal antibody 23H9, with known LAC properties, delayed the lag time and reduced the peak height during thrombin generation induction in normal plasma dose-dependently (0-150 microg/ml). At variance, LAC patient 1:1 plasma mixtures manifested variable lag time prolongations and/or peak height reductions. Coupling these two most informative thrombin generation parameters in a peak height/lag time ratio, and upon normalization versus the normal plasma pool, this ratio distributed normally and was reduced in the plasma mixtures, for 59/60 known LAC plasmas. The normalized peak height/lag time ratio correlated well with the normalized dilute prothrombin time, diluted Russell's viper venom time and silica clotting time, measured in 1:1 plasma mixtures (correlation coefficients 0.59-0.72). The anticoagulant effects of activated protein C (0-7.5 nM) or 23H9 (0-150 microg/ml), spiked in the 1:1 LAC plasma mixtures were reduced for the majority of patients, compatible with functional competition between patient LAC and activated protein C and LAC and 23H9, respectively. Hence, the normalized thrombin generation-derived peak height/lag time ratio identifies LAC in plasma with high sensitivity in a single assay, irrespective of the patient's treatment with oral anticoagulants.
More Related Videos
Related Concept Videos
Peripheral Arterial Disease II: Clinical Manifestations and Diagnostic Evaluation
Automated Microbial Diagnostics