Development of a high performance liquid chromatography method for quantification of PAC-1 in rat plasma
Li-Na Fang1, Xiao-Hui Chen, Zhen Song
1Shenyang Pharmaceutical University, Shenyang 110016, People's Republic of China.
Abstract:
A sensitive and specific high performance liquid chromatography method with UV detection was developed and validated for the determination of PAC-1 in rat plasma. After extraction with ethyl acetate, the chromatographic separation was carried out on a Diamonsil C(18) column (150mmx4.6mm i.d., 5microm particle size, Zhonghuida) protected by a ODS guard column (10mmx4.6mm i.d., 5microm particle size), using acetonitrile-methanol-phosphate buffer (pH 3.0, 30mM) (31:3:66, v/v/v) as mobile phase at a flow rate of 1.0mL/min, and wavelength of the UV detector was set at 281nm. No interference from any endogenous substances was observed during the elution of PAC-1 and internal standard (IS, indapamide). The calibration curve was linear over the range of 0.05-20microg/mL (r>0.99). The lower limit of quantification was evaluated to be 50ng/mL. The method was successfully applied to the pharmacokinetic study of PAC-1 after intravenous and oral administration in rats, respectively.