Related Experiment Video
Updated: Jun 26, 2026

Characterization and Application of Passive Samplers for Monitoring of Pesticides in Water
Published on: August 3, 2016
Detection of organophosphate pesticides using a prototype liquid crystal monitor
John L Adgate1, Andrea Barteková, Peter C Raynor
1Division of Environmental Health Sciences School of Public Health, University of Minnesota, MMC 807, Room 1260 Mayo, 420 Delaware Street, S.E, Minneapolis, MN 55455-0392, USA. jadgate@umn.edu
Abstract:
The precision and accuracy of a prototype wearable liquid crystal monitor (LCM) for the measurement of airborne organophosphate pesticide concentrations was explored in a series of laboratory experiments. LCM response to vapor-phase and aerosol diazinon was compared to concentrations obtained using a standard reference method (NIOSH 5600) at concentrations ranging from approximately 8 to 108 ppb (parts per billion) over durations of 2 to 80 hours. Temperature ( approximately 25, 30, and 35 degrees C) and relative humidity (15, 50, and 85%) were varied to estimate the effect of these factors on LCM performance. The LCM response to vapor phase pesticide exposure was linear for concentrations in the range of 8-20 ppb. At exposure concentrations above approximately 20 ppb, however, there was a decline in monitor response and measurement precision. Elevated temperatures improved diazinon vapor-only measurement precision, while increased relative humidity reduced LCM response at the extremes of tested temperatures. Compared to vapor-only exposures, the LCM was less sensitive to diazinon aerosol concentrations, but displayed reasonable precision over a relatively large range of exposures (29 to 1190 ppb-hr). Further efforts to characterize temperature and humidity effects and improve low-end sensitivity would likely provide a portable personal exposure monitor or environmental sensor for this widely used class of pesticides.
Related Concept Videos
High-Performance Liquid Chromatography: Types of Detectors
Gas Chromatography: Types of Detectors-II