Field comparison of passive air samplers with reference monitors for ambient volatile organic compounds and nitrogen
Shaibal Mukerjee1, Karen D Oliver, Robert L Seila
1National Exposure Research Laboratory E205-02, U.S. Environmental Protection Agency, Research Triangle Park, NC 27711, USA.
Abstract:
This study evaluates performance of nitrogen dioxide (NO2) and volatile organic compound (VOC) passive samplers with corresponding reference monitors at two sites in the Detroit, Michigan area during the summer of 2005. Ogawa passive NO2 samplers and custom-made, re-useable Perkin-Elmer (PE) tubes with Carbopack X sorbent for VOCs were deployed under week-long sampling periods for six weeks. Precise results (5% relative standard deviation, RSD) were found for NO2 measurements from collocated Ogawa samplers. Reproducibility was also good for duplicate PE tubes for benzene, toluene, ethylbenzene, and xylene isomers (BTEX species, all < or = 6% RSD). As seen in previous studies, comparison of Ogawa NO2 samplers with reference chemiluminescence measurements suggested good agreement. Generally good agreement was also found between the PE tubes and reference methods for BTEX species.
More Related Videos
09:33Visualizing Field Data Collection Procedures of Exposure and Biomarker Assessments for the Household Air Pollution Intervention Network Trial in India
Published on: December 23, 2022
07:14Automated, High-resolution Mobile Collection System for the Nitrogen Isotopic Analysis of NOx
Published on: December 20, 2016