A new method for calibration and response measurement of a scintillation detector using radioisotope sources
Hidehito Nakamura1, Hiroyasu Ejiri, Hisashi Kitamura
1National Institute of Radiological Sciences, Chiba, Japan. hidehito@nirs.go.jp
Abstract:
A new method for calibration was developed that takes the energy deposited by radioisotope sources into account; it was validated with a plastic scintillation plate using a (207)Bi source. The energy deposits of the conversion electrons in the (207)Bi source led to an energy difference of a few keV for the 976 keV monoenergetic peak. Calibration with high accuracy is essential to estimate radiation doses as well as to evaluate detector performance. The method will be available not only for use in scintillation detectors but also for other detectors.
Related Concept Videos
Imaging Studies II: Positron Emission Tomography and Scintigraphy
Fundamental Principles of PET
Radioactive Decay and Radiometric Dating
Isotopes and Radioisotopes
An isotope containing more...
Atomic Absorption Spectroscopy: Instrumentation
The atomizer used in AAS can be either a flame atomizer or an...
Atomic Absorption Spectroscopy: Radiation and Light Sources
Two common narrow-range 'line' sources used in AAS are hollow-cathode lamps (HCLs) and...
Atomic Emission Spectroscopy: Instrumentation

![Automated Radiochemical Synthesis of [18F]3F4AP: A Novel PET Tracer for Imaging Demyelinating Diseases](/_next/image?url=https%3A%2F%2Fcloudfront.jove.com%2FCDNSource%2Fteasers%2F55537.jpg&w=3840&q=50)