A New Single Chamber Implantable Defibrillator with Atrial Sensing: A Practical Demonstration of Sensing and Ease of Implantation
Published on: February 28, 2012
Atrial fibrillation therapy in patients with a CRT defibrillator with wireless telemetry
George H Crossley1, Kazutaka Aonuma, Charles Haffajee
1St. Thomas Research Institute, and Division of Cardiology, University of Tennessee, College of Medicine, Nashville, Tennessee 37203, USA. gcrossley@saintthomasheart.com
Insights
Atrial shock therapy effectively treated atrial fibrillation (AF) in 98.2% of patients receiving a cardiac resynchronization therapy defibrillator (CRT-D). A new wireless telemetry system also demonstrated safe and effective performance in this patient population.
Area of Science:
- Cardiology
- Electrophysiology
- Medical Device Technology
Background:
- Atrial fibrillation (AF) significantly increases morbidity and mortality, particularly in patients with congestive heart failure.
- Cardiac resynchronization therapy defibrillators (CRT-D) are crucial for managing heart failure patients.
- Effective AF management is essential for improving outcomes in CRT-D recipients.
Purpose of the Study:
- To assess the efficacy of atrial shock therapy in CRT-D patients.
- To evaluate the safety and performance of a new CRT-D device.
- To determine the effectiveness of a novel wireless telemetry system for CRT-D.
Main Methods:
- Prospective, international, multicenter study involving 282 patients indicated for CRT-D.
- Atrial shock therapy tested for both spontaneous and induced AF episodes.
- Evaluation of the Medtronic Conexus wireless telemetry system's performance and safety.
Main Results:
- Atrial shock therapy achieved 98.2% efficacy in treating 171 AF episodes.
- Successful cardioversion rates varied by energy level (12 J, 24 J, 35 J).
- The wireless telemetry system demonstrated no loss of signal or adverse events across 1,999 uses.
Conclusions:
- Atrial shock therapy is highly effective (98.2%) for AF in patients meeting CRT-D indications.
- The evaluated wireless telemetry system is safe and performs effectively.
- These findings support the integrated use of atrial shock and wireless telemetry in CRT-D devices.
Background:
Atrial fibrillation (AF) is a major cause of morbidity and mortality, especially in patients with congestive heart failure.
Objectives:
The purposes of this international, prospective multicenter study were to evaluate the efficacy of atrial shock therapy in patients with a cardiac resynchronization therapy defibrillator (CRT-D) and to evaluate the safety of the new CRT-D. The effectiveness of a new wireless telemetry system was also evaluated.
Methods:
A total of 282 patients, without permanent AF, who had indications for a CRT-D were included. Atrial shock therapy was tested on both spontaneous and induced AF episodes. The effectiveness of the Medtronic wireless telemetry system (Conexus; Medtronic Inc., Minneapolis, MN, USA) was also tested. Secondary endpoints included the heart failure Clinical Composite Response, system performance evaluation, and adverse event summary.
Results:
Atrial shock therapy was successful in 168 of 171 episodes (98.2%). Of these, 138 episodes were induced and 33 were spontaneous. Successful cardioversion occurred in 137 of the 138 induced-AF episodes (86.1% with 12 joule (J), 13.1% with 24 J, and 0.7% with 35 J). During the first 3 months of implant, there were 43 system-related complications in 37 subjects out of 278 subjects. There were 1,999 Conexus telemetry uses recorded during this study. This includes 282 uses during the implant procedure. There were no cases of complete loss of telemetry or any adverse events reported using this system.
Conclusion:
We achieved an atrial shock efficacy of 98.2% in patients who met standard CRT-D indications. The wireless telemetry performed well with no reported unanticipated adverse device effects.
Related Concept Videos
ECG Interpretation of Arrhythmias II: Atrial, Junctional and Ventricular Arrhythmias
Dysrhythmias VI: Management of Dysrhythmias
Cardiopulmonary Resuscitation III: AED Use
Pulse rhythm
Conversely, an irregular pulse pattern is termed dysrhythmia, stemming from disruptions in cardiac muscle...
Disturbances in Heart Rhythm
Arrhythmias are categorized by their speed, rhythm, and origin. A slow heart...
Heart Failure VI: Adjunct Therapies