Strategy to eliminate inappropriate shocks secondary to T-wave oversensing in a biventricular ICD
Jonathan S Silver1, Mary Ellen Gray, Roy M John
1Lahey Clinic Medical Center, Division of Cardiovascular Medicine, Burlington, Massachusetts 01805, USA.
Abstract:
T-wave oversensing represents a common cause of inappropriate shocks in patients with implanted cardiac defibrillators. This case report demonstrates a strategy of device programming using V-V pace delay (sequential rather than simultaneous biventricular pacing) to eliminate T-wave oversensing without decreasing sensitivity to detect true tachyarrhythmia.