Isoelectric PVCs diagnosed with a monitor with simultaneous ECG, and ICD intracardiac electrogram and marker channels
David L Scher1, David C Man, Thomas Edery
1Division of Cardiac Electrophysiology, Associated Cardiologists, Harrisburg, Pennsylvania 17112, USA. dlsmd@comcast.net
Abstract:
A 78-year-old woman with a remote history of chronic atrial fibrillation and catheter ablation of the atrioventricular node with a permanent ventricular pacemaker sustained a syncopal episode. Interrogation of her pacemaker revealed electrograms of ventricular fibrillation with a duration of 26 seconds. The patient underwent upgrade to an implantable cardioverter defibrillator (ICD). Postimplant, on telemetry, pauses were seen without pacing artifact, suspicious for oversensing of something by the ICD. Interrogation and threshold testing of the ICD established that both sensing (of her junctional escape rhythm) and pacing thresholds were excellent. A Holter monitor was utilized to evaluate the patient's rhythm and device function. It incorporated simultaneous tracings from multiple surface electrocardiogram leads and both intracardiac electrograms and marker channel recordings from the ICD. The causes for the pauses were found to be isoelectric premature ventricular complexes not seen on the telemetry tracings.
Related Concept Videos
Pulse rhythm
Conversely, an irregular pulse pattern is termed dysrhythmia, stemming from disruptions in cardiac muscle...
Electrocardiogram
Three major waveforms are present in a typical ECG recording: the P wave, the QRS complex, and the T...
Electrocardiogram Fundamentals
An electrocardiogram (ECG) is a diagnostic tool for identifying cardiac conditions such as arrhythmias, conduction abnormalities, and myocardial ischemia.
Definition
An electrocardiogram (ECG) visualizes the heart's electrical activity by tracing the electrical movement associated with each heartbeat on a graph or monitor. As the heart beats, an electrical wave passes through it, correlating with the cardiac cycle events.
Parts of an ECG
An ECG utilizes electrodes on the skin to...
Holter Monitor: 24-Hour Monitoring
Dysrhythmias V: Evaluating Dysrhythmias
ECG Interpretation of Arrhythmias II: Atrial, Junctional and Ventricular Arrhythmias