Monoclonal antibody-based discrepancies between two-site immunometric tests for lutropin
K Pettersson1, Y Q Ding, I Huhtaniemi
1Department of Physiology, University of Turku, Finland.
Clinical Chemistry
|October 1, 1991
Summary
Monoclonal antibodies targeting intact lutropin (LH) can lead to inaccurate, low measurements in serum assays. Researchers should avoid these antibodies to ensure reliable LH testing and bioactivity monitoring.
Area of Science:
- Clinical Chemistry
- Endocrinology
- Immunodiagnostics
Background:
- Accurate measurement of lutropin (LH) is crucial for clinical diagnostics and research.
- Two-site immunometric assays are commonly used for LH quantification.
- Monoclonal antibodies targeting specific LH epitopes can influence assay performance.
Purpose of the Study:
- To compare the performance of five two-site immunometric assays for serum LH measurement.
- To evaluate the impact of monoclonal antibodies targeting the intact LH dimer on assay results.
- To identify potential discrepancies between immunoreactivity and bioactivity of LH.
Main Methods:
- Comparison of three commercially available and two in-house two-site immunometric assays for LH.
- Analysis of 83 serum samples, with a focus on 19 samples showing low LH concentrations.
- Assessment of LH bioactivity in five serum samples using in vitro methods.
Main Results:
- Four of the five assays, utilizing monoclonal antibodies against the intact LH dimer, reported significantly lower LH concentrations compared to a reference method.
- Two assays failed to detect significant LH immunoreactivity in five serum samples.
- The low immunoreactivity in these samples did not correlate with their normal in vitro LH bioactivity.
Conclusions:
- Monoclonal antibodies specific for the intact LH dimer can lead to underestimation of LH levels.
- These assays may fail to detect bioactive LH, causing confusion in clinical and research settings.
- The study recommends avoiding monoclonal antibodies specific for the intact LH dimer in LH assays.