Methodology of QT-interval measurement in the modular ECG analysis system (MEANS)
Jan A Kors1, Gerard van Herpen
1Department of Medical Informatics, Erasmus University Medical Center, The Netherlands. j.kors@erasmusmc.nl
Background:
QT prolongation as can be induced by drugs, signals the risk of life-threatening arrhythmias. The methodology of QT measurement in the modular ECG analysis system (MEANS) is described.
Methods:
In the simultaneously recorded leads of the standard 12-lead electrocardiogram (ECG), the QRS complexes are detected by a spatial velocity function. They are typed as dominant or nondominant, and a representative complex per lead is obtained by averaging over the dominant complexes. QRS onset and T end are determined by a template technique, and QT is measured. MEANS performance was evaluated on the 125 ECGs of the common standards for quantitative electrocardiography (CSE) multilead database, of which the waveform boundaries have been released.
Results:
MEANS detected correctly all 1445 complexes of the CSE library, with one false-positive detection due to a sudden baseline jump. All dominant complexes were correctly typed. The average of the differences between MEANS and reference was less than 2 ms (=1 sample) for both QRS onset and T end, and 2.1 ms for QT duration. The standard deviation of the differences was 3.8, 8.4, and 10.4 ms, respectively.
Conclusions:
A standard deviation of 10.4 ms for QT measurement seems large when related to the regulatory requirement that a prolongation as small as 5 ms should be detected. However, QT variabilities as encountered in different individuals will be larger than when measured in one individual during pharmacological intervention. Finally, if the U wave is part of the total repolarization, then T and U form a continuum and the end of T becomes questionable.
Related Concept Videos
Electrocardiogram
Three major waveforms are present in a typical ECG recording: the P wave, the QRS complex, and the T...
ECG Interpretation of Rhythms
Components of the Electrocardiogram
The primary components of a normal ECG waveform in Normal sinus rhythm(NSR) include the P wave, PR interval, QRS complex, ST segment, T wave, and occasionally a U wave.
ECG waveforms are divided by vertical and horizontal lines at standard intervals.
The horizontal axis measures time and rate, and the vertical axis measures amplitude or voltage. When...
Electrocardiogram Fundamentals
An electrocardiogram (ECG) is a diagnostic tool for identifying cardiac conditions such as arrhythmias, conduction abnormalities, and myocardial ischemia.
Definition
An electrocardiogram (ECG) visualizes the heart's electrical activity by tracing the electrical movement associated with each heartbeat on a graph or monitor. As the heart beats, an electrical wave passes through it, correlating with the cardiac cycle events.
Parts of an ECG
An ECG utilizes electrodes on the skin to...
Correlation between ECG and Cardiac Cycle
A cardiac action potential originates in the SA node and spreads throughout the atria and the AV node in approximately 0.03 seconds. This results in the P wave in an ECG and triggers atrial contraction. The action potential is then briefly slowed at the AV node, allowing the atria to contract and fill the ventricles with blood before...
ECG Interpretation of Arrhythmias II: Atrial, Junctional and Ventricular Arrhythmias
Dysrhythmias IV: Characteristics of Bradyarrhythmias