Protocol for Relative Hydrodynamic Assessment of Tri-leaflet Polymer Valves
Published on: October 17, 2013
Two-dimensional simulation of flow and platelet dynamics in the hinge region of a mechanical heart valve
V Govindarajan1, H S Udaykumar, K B Chandran
1Department of Biomedical Engineering, College of Engineering, The University of lowa, 1402 SC, lowa City, IA 52242, USA.
Abstract:
The hinge region of a mechanical bileaflet valve is implicated in blood damage and initiation of thrombus formation. Detailed fluid dynamic analysis in the complex geometry of the hinge region during the closing phase of the bileaflet valve is the focus of this study to understand the effect of fluid-induced stresses on the activation of platelets. A fixed-grid Cartesian mesh flow solver is used to simulate the blood flow through a two-dimensional geometry of the hinge region of a bileaflet mechanical valve. Use of local mesh refinement algorithm provides mesh adaptation based on the gradients of flow in the constricted geometry of the hinge. Leaflet motion is specified from the fluid-structure interaction analysis of the leaflet dynamics during the closing phase from a previous study, which focused on the fluid mechanics at the gap between the leaflet edges and the valve housing. A Lagrangian particle tracking method is used to model and track the platelets and to compute the magnitude of the shear stress on the platelets as they pass through the hinge region. Results show that there is a boundary layer separation in the gaps between the leaflet ear and the constricted hinge geometry. Separated shear layers roll up into vortical structures that lead to high residence times combined with exposure to high-shear stresses for particles in the hinge region. Particles are preferentially entrained into this recirculation zone, presenting the possibility of platelet activation, aggregation, and initiation of thrombi.
Related Concept Videos
Steady, Laminar Flow Between Parallel Plates
Fluid Pressure over Curved Plate of Constant Width
Fluid Pressure over Flat Plate of Variable Width
The pressure distribution on the plate can be calculated by determining the force that acts on a differential area strip of the plate. Thus, the magnitude of the force is equal to the...
Heart Valves
The AV valves prevent the backflow of blood from the ventricles to the atria during ventricular contraction. These valves function with the assistance of the chordae tendineae and papillary muscles. When the ventricles are relaxed, the chordae tendineae are slack, allowing blood to flow from the atria into the...
Fluid Pressure over Flat Plate of Constant Width
The resultant force...