Comparison of tungsten coil electrothermal vaporization and thermospray sample introduction methods for flame furnace atomic absorption spectrometry
Peng Wu1, Yunchang Zhang, Rui Liu
1Analytical & Testing Center, Sichuan University, Chengdu, Sichuan 610064, China.
Abstract:
Flame furnace atomic absorption spectrometry (FF-AAS) is a newly developed flame atomic absorption spectrometric technique based on arranging a flame furnace onto the top of the flame burner head. In this fundamental investigation, 25 elements were carefully tested by using either thermospray FF-AAS or tungsten coil electrothermal vaporization FF-AAS, of which 15 volatile and semi-volatile elements (Cd, Tl, Ag, Pb, Zn, Hg, Cu, Sb, Bi, Te, In, As, Se, Sn and Au) exhibited better limits of detection compared to those by conventional FAAS; however, non-volatile or refractory elements (Fe, Co, Ni, Cr, Mn, Pd, Pt, Al, Be and V) showed inferior sensitivities by the proposed methods.
Related Concept Videos
Atomic Absorption Spectroscopy: Atomization Methods
Flame Photometry: Overview
Atomic Emission Spectroscopy: Instrumentation
Atomic Absorption Spectroscopy: Lab
Solutions containing organic solvents, such as low-molecular-mass alcohols, esters, or ketones, enhance absorbances by increasing nebulizer...
Atomic Spectroscopy: Effects of Temperature
At thermal equilibrium, the relative populations of excited and ground state atoms can be estimated using the Maxwell–Boltzmann distribution. For example, an increase in temperature from...