Related Experiment Video
Updated: Jun 26, 2026

A Computational Modeling Approach to Investigate the Influence of Hyperthermia on the Tumor Microenvironment
Published on: December 1, 2023
Heating properties of re-entrant resonant cavity applicator for brain tumor with simple head model
T Hirashima1, K Kato, Y Shindo
1Department of Mechanical Engineering Informatics, Meiji University, Kawasaki, Japan.
Abstract:
We have already confirmed the effectiveness of the re-entrant resonant cavity applicator system with non-invasive experiments of heating cylindrical agar phantoms and computer simulations. This paper discusses the heating properties of the developed heating system with a human head model made of agar for brain tumor hyperthermia treatment. First, we present the results of heating a uniform agar head model with the developed heating system. In the experiments, the temperature rise at the center of the agar was about 8 degrees C, it was found that the center of the agar is heated to maximum temperature non-invasively. Second, we present the results of heating a non-uniform agar head model having an oral cavity and a nasal cavity. We found that the center of the agar can be heated to maximum temperature as well as uniform agar head model. From these results, it is confirmed that the possibility of effective hyperthermia for various types of deep-seated brain tumors.