Variations in non-invasive radial artery pulse waveform at different measuring positions
Jeon Lee1, Young-Ju Jeon, Hae-Jung Lee
1Research Institute of Medical Instruments and Rehabilitation Engineering, Medical Industry Techno Tower, 234 Maeji, Heungup, Wonju, Kangwon, 220-710, South Korea. leejeon@kiom.re.kr
Abstract:
Noninvasive radial artery pulse wave has been widely not only for the pulse wave analysis (PWA) itself but also for assessment of arterial stiffness with estimated aortic pulse wave from peripheral pulse wave. In this study, we tested the differences between the extracted parameters, including AIx, from radial artery pulse waves measured at three different positions preliminarily. A 30 years old female involved in this experiment and this subject has no evidence about cardiovascular diseases. Radial pulse waves were recorded at three different positions on left lower arm ten times for each position. These positions are widely used in pulse diagnosis of traditional medicine.As results of repeated measures ANOVA analysis, 9 parameters were found which had significantly differences in between measuring positions (P<0.05). These parameters wereP_AIx, PP_Amp, A_T1, A_Tr, AIx, AIx_@HR75, ESP, MP_Sysand MP_Dia. And, by Spearman correlation test at P2, it was observed that P_AIx (r =0.894), PP_Amp (r=-0.765),AIx_@HR75 (r=0.929), ESP (r=0.893) and MP_Dia (r=0.792)were highly correlated with AIx. Elaborately, the correlation coefficients between each parameter were high in similar degrees (r>0.70).Further studies on what make differences in these parameters between measuring positions are need. And, ultimately, we need to determine the best measuring process, including measuring position, hold-down pressure, signal quality validation and so on,so to achieve the optimal waveform which represents subject's health condition for both western medicine and traditional medicine.
Related Concept Videos
Assessment of radial pulse
The radial pulse, located at the wrist, is often the preferred site for assessing peripheral pulse because of its accessibility and dependability. The process of determining the radial pulse involves several steps:
Assessment of apical radial pulse
The A-R pulse assessment involves simultaneous evaluation of the apical and radial pulses. When the apical and radial pulse rates vary, this assessment helps identify a pulse deficit.
Pre-Procedural Preparation
Pulse
The pulse serves as a clinical indicator...
Special considerations while measuring pulse
Sites for measuring blood pressure
The Brachial Artery: Primary Site for Blood Pressure Measurement
Pulse Assessment Sites