Comparison of reflected green light and infrared photoplethysmography
1Graduate School of Engineering, Chiba University, Chiba, Japan. y.maeda@graduate.chiba-u.jp
Abstract:
We evaluated the accuracy of pulse rate measurements obtained by reflected green light photoplethysmography (PPG) compared to reflected infrared light photoplethysmography and ECG. The wavelengths of the green and infrared light were 525 and 880 nm, respectively, and experiments were performed at 25 degrees C and at a skin temperature below 15 degrees C. The pulse rate obtained from reflected green light PPG was compared with the ECG RR interval and the pulse rate from reflected infrared PPG. The results indicated a stronger correlation between green PPG and ECG results at both temperatures. These results suggested that reflected green light PPG had an advantage over reflected infrared PPG, especially at temperatures below 15 degrees C.