Related Experiment Video
Updated: Jun 26, 2026

Performing Repeated Intraoperative Impedance Telemetry Measurements during Cochlear Implantation
Published on: August 4, 2023
Radiofrequency dosimetry in subjects implanted with metallic straight wires: a numerical study
E Mattei1, G Calcagnini, F Censi
1Italian National Institute of Health, Rome, Italy. eugenio.mattei@iss.i
Abstract:
A numerical study to investigate the effects of the exposure to electromagnetic fields (EMF) at 900 and 1800 MHz on biological tissues implanted with thin metallic structures has been carried out, using the finite difference time domain (FDTD) solution technique. The results of the model show that the presence of a metallic wire yields to a significant increase in the local specific energy absorption rate (SAR). The present standards and/or guidelines on safe exposures of humans to EMF does not cover persons with implanted devices and thus the threshold levels to define safe exposure conditions might not apply in presence of high SAR gradients, such as the ones generated by thin metallic implanted objects. However, exposure to EMF fields below the actual safe levels even in presence of thin conductive structures cause rather low temperature rises (1 degrees C).