Evaluation of Fluid Overload by Bioelectrical Impedance Vectorial Analysis
Published on: August 17, 2022
Usefulness of monitoring congestive heart failure by multiple impedance vectors
Dorin Panescu1, Mihir Naware, Jeff Siou
1St. Jude Medical, Sunnyvale, CA, USA.
Introduction:
We investigated trends in intrathoracic impedance measured between multiple implanted electrodes for monitoring pulmonary edema secondary to congestive heart failure (CHF) in an experimental model.
Methods:
Biventricular ICDs were implanted in 16 dogs and 5 sheep. Continuous RV pacing (230- 250 bpm) was applied over several weeks. Meanwhile, impedance was measured every hour along 4 intrathoracic and 2 intracardiac vectors. Four cardiogenic impedance vectors were also monitored. Cardiac function was assessed biweekly by catheterization and echocardiography. Left atrial (LA) pressure was measured daily by an implanted LA pressure sensor.
Results:
All animals developed CHF after 2-4 weeks of pacing as evidenced by changes in function (EF, 52 vs. 34%; LV end-diastolic volume, 65 vs. 97 ml; LV end-diastolic pressure, 7 vs. 16 mmHg; LA volume, 17 vs. 33 ml; LA pressure, 7 vs. 26 mmHg), clinical symptoms, or autopsy. Steady state impedance decreased during CHF: LV-Can, 17+/-9%; LV-RV, 15+/-8%; LV-RA, 13+/-6%; RV-Can, 13+/-8%; RVcoil-Can, 8+/-6%; RA-Can, 6+/-6%. Change in LV-Can impedance was greater than that of RA-Can, RV-Can, and RVcoil-Can (P0.05). LV-Can impedance correlated well with LA pressure (r(2)=0.73), while RV-Can and RVcoil-Can were weakly correlated (r(2)=0.43 and r(2)=0.52, respectively). Changes in LV-RV and LV-RA impedances were also larger than those of RVcoil-Can and RA-Can (P0.05). Meanwhile, all impedances were associated with circadian variability at baseline (5+/-2%) which diminished during CHF (2+/-1%); P=0.02. Furthermore, significant variations were observed in cardiogenic impedances during progression into CHF as evidenced by reduced peak-to-peak amplitude and increased fractionation of the signals.
Conclusions:
All impedance vectors decreased during CHF. Impedance measurement employing left heart sensors correlated well wit- - h LA pressure, and may improve detection of CHF onset compared to sensing by RA or RV leads alone. This approach has important clinical implications for managing heart failure patients in the ambulatory setting.
More Related Videos
10:21Evaluation of Hydration Status by Bioelectrical Impedance Vector Analysis in Patients with Ischemic Heart Disease Undergoing Exercise Stress Test
Published on: September 22, 2023
04:05Clinical Application of Phase Angle and BIVA Z-Score Analyses in Patients Admitted to an Emergency Department with Acute Heart Failure
Published on: June 30, 2023
Related Concept Videos
Pulse rhythm
Conversely, an irregular pulse pattern is termed dysrhythmia, stemming from disruptions in cardiac muscle...
Holter Monitor: 24-Hour Monitoring
Equipments Used To Measure Blood Pressure
This invasive approach involves cannulating a peripheral artery. During each cardiac contraction, pressure generates mechanical motion within the catheter, transmitted through rigid, fluid-filled tubing to a transducer. This transducer converts mechanical motion into electrical signals displayed as waveforms on a monitor. An automatic flushing system prevents blood backflow. Due to the potential risk of unexpected arterial blood loss, this method is primarily used in intensive...
Dysrhythmias V: Evaluating Dysrhythmias