Scanning Electron Microscopic Evaluation of Surface Defects of Remover Retreatment File After Single and Multiple Uses
Published on: October 11, 2024
Defects in nickel-titanium instruments after clinical use. Part 4: an electropolished instrument
Ya Shen1, Earl Winestock, Gary Shun-pan Cheung
1Division of Endodontics, Department of Oral Biological & Medical Sciences, Faculty of Dentistry, University of British Columbia, Vancouver, Canada.
Abstract:
This study aimed to examine the type and location of defects observed in one brand of electropolished nickel-titanium instruments after routine clinical use. A total of 414 RaCe instruments with structural deformations, discarded from one endodontic clinic during a period of 24 months, were analyzed. The lateral and fracture surfaces of those separated instruments were examined by scanning; the locations of the fractures were recorded. Of all defective instruments, 388 had an area of structural change, and 26 were fractured. The distance from the tip of the instrument to the end of the deformed area was significantly longer in .02 taper instruments than in other tapers (P < 0.01), whereas in .04 instruments the distorted area ended closer to the tip (P < 0.05). In 31% of all fractured instruments, the fragment was 7-8 mm long, more frequently in .04 taper than in others. Most of the fractures were found adjacent to the spiraled sector. Shear fracture was diagnosed for 22 (85%) of all those fractured, whereas fatigue occurred in only 4 instruments (15%) (P < 0.05). There was a marked increase in the amount of defects after the seventh (45.4%) and eighth (100%) use. The results of this study indicated that RaCe instruments commonly exhibited unwinding defects after repeated use, and the most common cause of failure was shear failure.
More Related Videos
06:48Rapid Formation and Testing of Self-expanding NiTi Frames with a Small Form Factor Suitable for Minimally Invasive Implants
Published on: March 7, 2025
06:36Effects of Mechanical Methods Used in Peri-implantitis Treatment on Implant Surface Decontamination and Roughness
Published on: March 14, 2025
Related Concept Videos
Imperfections in Crystal Structure: Non-Stoichiometric Defects
Precipitation Gravimetry
In determining nickel by gravimetric analysis, a precipitant of ethanolic dimethylglyoxime is added to a hot nickel salt solution. This is quickly followed by the dropwise addition of dilute ammonia solution until precipitation occurs. A...