Accuracy in Dental Medicine, A New Way to Measure Trueness and Precision
Published on: April 29, 2014
Accuracy of electronic apex locater-controlled handpieces
Jonathan Barthelemy1, Ladislav Gregor, Ivo Krejci
1Division of Cariology and Endodontology, School of Dental Medicine, University of Geneva, Geneva, Switzerland.
Abstract:
An in vitro model was used to test a hypothesis that the manual (apex locator) and motor-driven (apex locator-controlled handpiece) operating modes of 3 newly developed apex-locator-controlled handpiece devices (Dentaport ZX, Endomaster, XSmart Dual) give the same working length. The depth of penetration of the file into the root canal was measured using a digital micrometer and the distance of the tip of the file relative to the major root foramen was determined by scanning electron microscopy (SEM). In the manual mode, the XSmart Dual device reported significantly shorter working lengths than the Dentaport ZX or the Endomaster devices. In the motor-driven mode, the XSmart Dual device reported significantly longer working lengths than the Dentaport ZX but not the Endomaster. Most instruments driven by the handpieces were confined within the root canal and differences in working lengths between manual and motor-driven modes were small for all devices (tenths of millimeters). We concluded that although the motor-driven mode of these devices appeared to be clinically safe, measurements obtained in manual and motor-driven operating modes are not equivalent.
Related Concept Videos
Electronic Distance Measuring Instruments
Accuracy and Precision
Assessment of apical radial pulse
The A-R pulse assessment involves simultaneous evaluation of the apical and radial pulses. When the apical and radial pulse rates vary, this assessment helps identify a pulse deficit.
Pre-Procedural Preparation
Assessment of apical pulse
Assessing the apical pulse is a critical nursing procedure, particularly indicated for: