Objective:
The purpose of this study was to develop a high-resolution, near-field-optimized 14-MHz, 24-element broad-bandwidth forward-looking array for integration on a steerable 9F electrophysiology (EP) catheter.
Methods:
Several generations of prototype imaging catheters with bidirectional steering, termed microlinear (ML), were built and tested as integrated catheter designs with EP sensing electrodes near the tip. The wide-bandwidth ultrasound array was mounted on the very tip, equipped with an aperture of only 1.2 by 1.58 mm. The array pulse echo performance was fully simulated, and its construction offered shielding from ablation noise. Both ex vivo and in vivo imaging with a porcine animal model were performed.
Results:
The array pulse echo performance was concordant with Krimholtz-Leedom-Matthaei model simulation. Three generations of prototype devices were tested in the right atrium and ventricle in 4 acute pig studies for the following characteristics: (1) image quality, (2) anatomic identification, (3) visualization of other catheter devices, and (4) for a mechanism for stabilization when imaging ablation. The ML catheter is capable of both low-artifact ablation imaging on a standard clinical imaging system and high-frame rate myocardial wall strain rate imaging for detecting changes in cardiac mechanics associated with ablation.
Conclusions:
The imaging resolution performance of this very small array device, together with its penetration beyond 2 cm, is excellent considering its very small array aperture. The forward-looking intracardiac catheter has been adapted to work easily on an existing commercial imaging platform with very minor software modifications.
More Related Videos
06:57Ablation of Ischemic Ventricular Tachycardia Using a Multipolar Catheter and 3-dimensional Mapping System for High-density Electro-anatomical Reconstruction
Published on: January 31, 2019
04:29Three-Dimensional Modeling of the Left Atrium and Pulmonary Veins with a Precise Intracardiac Echocardiography Approach
Published on: June 30, 2023
Related Concept Videos
Imaging Studies for Cardiovascular System II:Types of Echocardiography
Types of Echocardiography
Transthoracic Echocardiography (TTE)
TTE is the most common type of echocardiogram which involves placing a transducer on the patient's chest, emitting sound waves to create heart images. TTE is invaluable for evaluating the heart's size, structure, and motion, making it particularly useful for diagnosing...
Imaging Studies for Cardiovascular System I:Echocardiography
Indications: Echocardiography is utilized to diagnose heart failure, valve disorders, and myocardial infarction. It also assesses cardiac structures' size, shape, and motion, evaluates...