Limitations of 64-detector-row computed tomography coronary angiography: calcium and motion but not short experience
H Mir-Akbari1, J Ripsweden, J Jensen
1Division of Cardiology, Department of Internal Medicine, Karolinska Institute, Karolinska University Hospital, Stockholm, Sweden.
Background:
Recently, 64-detector-row computed tomography coronary angiography (CTA) has been introduced for the noninvasive diagnosis of coronary artery disease.
Purpose:
To evaluate the diagnostic capacity and limitations of a newly established CTA service.
Material And Methods:
In 101 outpatients with suspected coronary artery disease, 64-detector-row CTA (VCT Lightspeed 64; GE Healthcare, Milwaukee, Wisc., USA) was performed before invasive coronary angiography (ICA). The presence of >50% diameter coronary stenosis on CTA was rated by two radiologists recently trained in CTA, and separately by an experienced colleague. Diagnostic performance of CTA was calculated on segment, vessel, and patient levels, using ICA as a reference. Segments with a proximal reference diameter <2 mm or with stents were not analyzed.
Results:
In 51 of 101 patients and 121 of 1280 segments, ICA detected coronary stenosis. In 274 of 1280 (21%) segments, CTA had non-diagnostic image quality, the main reasons being severe calcifications (49%), motion artifacts associated with high or irregular heart rate (45%), and low contrast opacification (14%). Significantly more women (43%) had non-diagnostic scans compared to men (20%). A heart rate above 60 beats per minute was associated with significantly more non-diagnostic patients (38% vs. 18%). In the 1006 diagnostic segments, CTA had a sensitivity of 78%, specificity of 95%, positive predictive value (PPV) of 54%, and negative predictive value (NPV) of 98% for detecting significant coronary stenosis. In 29 patients, CTA was non-diagnostic. In the remaining 72 patients, sensitivity was 100%, specificity 65%, PPV 79%, and NPV 100%. The use of a more experienced CTA reader did not improve diagnostic performance.
Conclusion:
CTA had a very high negative predictive value, but the number of non-diagnostic scans was high, especially in women. The main limitations were motion artifacts and vessel calcifications, while short experience in CTA did not influence the interpretation.
More Related Videos
Related Concept Videos
Imaging Studies for Cardiovascular System V: CT
Imaging Studies for Cardiovascular System VI: Calcium -Scoring CT
Computed Tomography
The technique was invented in the 1970s and is based on the principle that as X-rays pass through the body, they are absorbed or reflected at different levels. In the technique, a patient lies on a motorized platform while a computerized axial tomography (CAT) scanner rotates...
Imaging Studies for Cardiovascular System III: X-Ray
Definition and Purpose
An X-ray, or radiograph, is a non-invasive method that uses ionizing radiation to take images of internal structures. It is mainly used in cardiac imaging to examine the heart, lungs, and major blood vessels, aiming to identify abnormalities in the heart's size, shape, and position, such as heart failure, congenital defects, and vascular...
Imaging Studies for Cardiovascular System IV: CMRI
Imaging Studies III: Computed Tomography