Cone beam computed tomography, a low-dose imaging technique in the postoperative assessment of cochlear implantation
Joana Ruivo1, Koen Mermuys, Klaus Bacher
1Hospital de Santa Maria, Lisbon, Portugal.
Background:
Cone beam computed tomography (CBCT) has become an extremely useful technique for dentomaxillofacial imaging because it provides clear images of highly contrasted structures. Previous studies evaluating the applicability of this technique in otologic imaging were very encouraging but were only performed in vitro on temporal bone specimens. The intracochlear positioning of the individual electrodes after cochlear implantation by means of CBCT has not yet been shown in vivo.
Objective:
We describe a protocol for in vivo postoperative imaging of cochlear implants by CBCT. Moreover, the effective dose was measured and compared with the effective dose used on 4- and 16-slice multislice computed tomography (MSCT) by using a RANDO-phantom.
Main Outcome Measure:
Developing a protocol for in vivo postoperative imaging of cochlear implants by CBCT.
Results:
CBCT provides high-resolution and almost artifact-free multiplanar reconstruction images allowing assessment of the precise intracochlear position of the electrode and visualization of each of the individual contacts. The calculated effective dose of the used CBCT and MSCT acquisitions is 80 musv for the CBCT, 3,600 musv for the 16-slice computed tomography, and 4,800 musv for the 4-slice computed tomography.
Conclusion:
These preliminary results suggests that, for in vivo postoperative evaluation of cochlear implants, CBCT can provide at least the same information as conventional radiography, digital radiograph, and MSCT but in a more comfortable and a much more safer way.
Related Concept Videos
Computed Tomography
The technique was invented in the 1970s and is based on the principle that as X-rays pass through the body, they are absorbed or reflected at different levels. In the technique, a patient lies on a motorized platform while a computerized axial tomography (CAT) scanner rotates...
Imaging Studies III: Computed Tomography
Imaging Studies I: CT and MRI
Description of the Procedures
Computed Tomography (CT) scan:
Computed Tomography (CT) scans use X-ray technology to generate detailed images of bones, organs, and tissues. During the scan, the patient lies on a moving table...
Imaging Studies for Cardiovascular System V: CT