Validation of bioanalytical LC-MS/MS assays: evaluation of matrix effects
Ann Van Eeckhaut1, Katrien Lanckmans, Sophie Sarre
1Vrije Universiteit Brussel, Department of Pharmaceutical Chemistry, Drug Analysis and Drug Information, Laarbeeklaan 103, 1090 Brussels, Belgium.
Abstract:
Liquid chromatography coupled to atmospheric pressure ionization tandem mass spectrometry is currently the method of choice for the quantitative determination of drugs in biological matrices. The advantages of this technique include high specificity, sensitivity and throughput. However, co-eluting matrix components, which are not observed in the chromatogram, can have a detrimental effect on the analysis, since they can cause ion suppression or enhancement of the analyte. The evaluation of matrix effects on the quantitative analysis of drugs in biological fluids is an important and sometimes overlooked aspect of assay validation. In this review, the influence of matrix effects on bioanalytical LC-MS/MS methods is discussed and illustrated with some examples. In addition, possible solutions to reduce or eliminate matrix effects are highlighted. A literature overview of validated LC-MS/MS methods published from January till June 2008 is also included. Although matrix effects are investigated in most papers, there is no consensus on how matrix effects should be evaluated during method validation. In addition, the definition of specificity should be changed for LC-MS/MS based methods.
More Related Videos
09:49Investigation of Microbial Cooperation via Imaging Mass Spectrometry Analysis of Bacterial Colonies Grown on Agar and in Tissue During Infection
Published on: November 18, 2022
08:56Detection of Regulated Ergot Alkaloids in Food Matrices by Liquid Chromatography-Trapped Ion Mobility Spectrometry-Time-of-Flight Mass Spectrometry
Published on: November 22, 2024
Related Concept Videos
Inductively Coupled Plasma-Mass Spectrometry (ICP-MS): Interferences
Matrix-Assisted Laser Desorption Ionization (MALDI)
Data Validation
Key parameters for method validation include: