Effect of radiofrequency energy on glenohumeral fluid temperature during shoulder arthroscopy
Christopher R Good1, Michael K Shindle, Matthew H Griffith
1Department of Orthopedic Surgery, Sports Medicine and Shoulder Service, The Hospital for Special Surgery, New York, NY, USA.
Background:
Reports of glenohumeral chondrolysis following arthroscopy have raised concern about the deleterious effects that thermal devices may have on articular cartilage. The purpose of this study was to investigate the effects of flow and duration of treatment with a thermal device on temperatures within cadaveric glenohumeral joint specimens. It was hypothesized that the use of a thermal device during surgery increases the temperature of fluid within the joint to >45 degrees C, which has been shown to cause chondrocyte death.
Methods:
Temperature was measured at four locations within ten cadaver shoulder joints. Eight heating trials were performed on each cadaver shoulder to test three variables: the method of heating (continuous or intermittent), the fluid-pump flow rate (no flow, 50% flow, or 100% flow), and the location of the radiofrequency probe (the radiofrequency energy was either applied directly to anterior capsular tissue in a paintbrush pattern or held adjacent to the glenoid without tissue contact).
Results:
Temperatures of >45 degrees C occurred in every trial. The average maximum temperatures in all no-flow conditions were significantly higher than those in the trials with flow. Higher temperatures were measured by the anterior probe in all trials. When the heating had been applied adjacent to the glenoid, without tissue contact, the time needed to cool to a safe temperature was significantly longer in the no-flow states (average, 140.5 seconds) than it was in the 50% flow states (average, 12.5 seconds) or the 100% flow states (average, 8.5 seconds).
Conclusions:
Use of a thermal probe during arthroscopy may cause joint fluid temperatures to reach levels high enough to cause chondrocyte death. Maintaining adequate fluid-pump flow rates may help to lower joint fluid temperatures and protect articular cartilage.
Related Concept Videos
Assessing Body Temperature - Axilla
Step 1: Perform hand hygiene and put on clean gloves to maintain infection control and prevent cross-contamination.
Step 2: Prepare the patient by explaining the procedure to ensure understanding and cooperation. Ensure privacy, expose the axilla, and inform the patient that minimal movement is crucial for an accurate reading.
Step 3: Adjust the patient’s clothing to expose only the axilla. It minimizes...
Assessing Body Temperature - Temporal Artery
Step 1: Perform hand hygiene and don a fresh pair of gloves to prevent cross-infection and ensure patient safety.
Step 2: Explain the procedure to the patient to establish trust. Clear communication establishes trust with the patient, ensures they understand what to expect, promotes cooperation, and enhances comfort during the procedure.
Step 3: Assess the patient's forehead...