Related Experiment Video
Updated: Jun 26, 2026

Frequency Mixing Magnetic Detection Scanner for Imaging Magnetic Particles in Planar Samples
Published on: June 9, 2016
Limitations of calculating field distributions and magnetic susceptibilities in MRI using a Fourier based method
Yu-Chung N Cheng1, Jaladhar Neelavalli, E Mark Haacke
1Department of Radiology, Wayne State University, Detriot, MI, USA.
Abstract:
A discrete Fourier based method for calculating field distributions and local magnetic susceptibility in MRI is carefully studied. Simulations suggest that the method based on discrete Green's functions in both 2D and 3D spaces has less error than the method based on continuous Green's functions. The 2D field calculations require the correction of the 'Lorentz disk', which is similar to the Lorentz sphere term in the 3D case. A standard least-squares fit is proposed for the extraction of susceptibility for a single object from MR images. Simulations and a phantom study confirm both the discrete method and the feasibility of the least-squares fit approach. Finding accurate susceptibility values of local structures in the brain from MR images may be possible with this approach in the future.
Related Concept Videos
Magnetic Fields
A magnetic field is defined by the force that a charged particle experiences...
Magnetic Resonance Imaging
Magnetic Field due to Moving Charges
Consider a point charge moving with a constant velocity. Like the electric field, the magnetic field at any point is directly proportional to the magnitude of the charge and inversely proportional to the square of the distance between the source point and the field point. However, unlike the electric field, the magnetic field is always perpendicular to the plane containing the line...
Induced Electric Fields: Applications
Magnetic Field Due to Two Straight Wires
Plane Electromagnetic Waves II