Related Experiment Video
Updated: Jun 26, 2026

Failure of Cleaning Verification in Pharmaceutical Industry Due to Uncleanliness of Stainless Steel Surface
Published on: August 11, 2017
Validated UPLC method for the fast and sensitive determination of steroid residues in support of cleaning validation
Szabolcs Fekete1, Jeno Fekete, Katalin Ganzler
1Formulation Development, Chemical Works of Gedeon Richter Plc, Budapest X, Gyömroi út 19-21, Hungary. fekete.szabolcs1@chello.hu
Abstract:
An ultra performance liquid chromatographic (UPLC) method was developed for simultaneous determination of seven steroid (dienogest, finasteride, gestodene, levonorgestrel, estradiol, ethinylestradiol, and norethisterone acetate) active pharmaceutical ingredient (API) residues. A new, generic method is presented, with which it is possible to verify the cleaning process of a steroid producing equipment line used for the production of various pharmaceuticals. The UPLC method was validated using an UPLC BEH C18 column with a particle size of 1.7 microm (50 mm x 2.1 mm) and acetonitrile-water (48:52, v/v) as mobile phase at a flow rate of 0.55 ml/min. Method development and method validation for cleaning control analysis are described. The rapid UPLC method is suitable for cleaning control assays within good manufacturing practices (GMP) of the pharmaceutical industry.