Effect of tube current on diagnostic image quality in paediatric cerebral multidetector CT images
K Ledenius1, M Gustavsson, S Johansson
1Department of Radiation Physics, Sahlgrenska Academy, University of Gothenburg, Sahlgrenska University Hospital, SE-413 45 Göteborg, Sweden. kerstin.ledenius@vgregion.se
Abstract:
The aim of this study was to investigate the effect of tube current on diagnostic image quality in paediatric cerebral multidetector CT (MDCT) images in order to identify the minimum radiation dose required to reproduce acceptable levels of different diagnostic image qualities. Original digital scanning data (raw data) were selected retrospectively from routine MDCT brain examinations of 25 paediatric patients. All examinations had been performed using axial scanning on an eight-slice MDCT (LightSpeed Ultra, GE Healthcare). Their ages ranged from newborn to 15 years. Quantum noise was added artificially to the raw data representing dose reductions equivalent to steps of 20 mA. Patient identification information was removed. Three experienced radiologists blindly and randomly assessed the resulting images from two different levels of the brain with regard to reproduction of structures and overall image quality. Final data were evaluated using the non-parametric statistical approach of inter-scale concordance. The minimum value of tube current-time product (mAs) required to reproduce an image of sufficient diagnostic quality was established in relation to the age of the patient. The corresponding CT dose index values by volume (CTDI(vol) (mGy)) were also established. In conclusion, acceptable reproduction of low-contrast structures was possible at CTDI(vol) values down to 20 mGy (patients 1-5 years old). For acceptable reproduction of high-contrast structures, CTDI(vol) values down to 10 mGy were considered possible (patients 1-5 years old). The original image quality for patients under 6 months of age (15 mGy) was found to be inadequate for acceptable reproduction of low-contrast structures.
Related Concept Videos
Computed Tomography
The technique was invented in the 1970s and is based on the principle that as X-rays pass through the body, they are absorbed or reflected at different levels. In the technique, a patient lies on a motorized platform while a computerized axial tomography (CAT) scanner rotates...
Imaging Studies for Cardiovascular System V: CT
Radiological Investigation I: X-ray and CT
Radiological Investigation II: MRI and Ventilation Perfusion Scan
Magnetic Resonance Imaging (MRI) and Ventilation Perfusion Scans are two radiological investigations that offer detailed diagnostic images of the body, particularly lung structures.
MRI
MRI uses magnetic fields and radiofrequency signals to distinguish between normal and abnormal tissues. This technology provides a more detailed diagnostic image than CT scans, enabling it to characterize pulmonary nodules, stage bronchogenic carcinoma, and evaluate inflammatory activity in...
Imaging Studies III: Computed Tomography
Imaging Studies I: CT and MRI
Description of the Procedures
Computed Tomography (CT) scan:
Computed Tomography (CT) scans use X-ray technology to generate detailed images of bones, organs, and tissues. During the scan, the patient lies on a moving table...