Left ventricular functional analysis with 16- and 64-row multidetector computed tomography: comparison with gated
Prachi P Agarwal1, Smita Patel, James Corbett
1Division of Cardiothoracic Radiology, Department of Radiology, University of Michigan Hospitals, Ann Arbor, MI 48109-0030, USA. prachia@med.umich.edu
Electrocardiogram-gated computed tomography (CT) and single-photon emission computed tomography (SPECT) show excellent correlation for left ventricular (LV) ejection fraction (EF). LV volumes are lower on SPECT compared to CT.
Area of Science:
- Cardiovascular Imaging
- Radiology
- Medical Physics
Background:
- Accurate assessment of left ventricular (LV) function is crucial in cardiovascular diagnostics.
- Electrocardiogram-gated computed tomography (CT) and single-photon emission computed tomography (SPECT) are non-invasive imaging modalities used for cardiac evaluation.
Purpose of the Study:
- To compare left ventricular (LV) ejection fraction (EF), end-diastolic volume (EDV), and end-systolic volume (ESV) measurements between electrocardiogram-gated CT and gated SPECT.
- To evaluate the diagnostic accuracy of CT in assessing LV volumes and function against SPECT.
Main Methods:
- Fifty-seven patients underwent both electrocardiogram-gated multidetector CT and SPECT within a 3-month period.
- LV EF, EDV, and ESV were calculated and compared between the two imaging modalities.
- Correlation coefficients and statistical significance were determined for all parameters.
Main Results:
- Excellent correlation was observed between LV EF derived from CT and SPECT (r = 0.81, P = 0.17).
- CT-derived EDV and ESV showed good correlation with SPECT (r = 0.88 and 0.94, respectively).
- LV volumes (EDV and ESV) were significantly lower on SPECT compared to CT (P = 0.0007 and P = 0.055, respectively).
- 64-row CT demonstrated superior correlation with SPECT for EF, EDV, and ESV compared to 16-row CT.
Conclusions:
- Gated CT, particularly 64-row multidetector CT, provides a good correlation with gated SPECT for LV EF, EDV, and ESV.
- LV volumes measured by SPECT tend to be lower than those measured by CT.
- CT is a reliable tool for assessing LV volumes and ejection fraction, comparable to SPECT.
More Related Videos
11:09High-Resolution Cardiac Positron Emission Tomography/Computed Tomography for Small Animals
Published on: December 16, 2022
11:13Quantification of Mouse Heart Left Ventricular Function, Myocardial Strain, and Hemodynamic Forces by Cardiovascular Magnetic Resonance Imaging
Published on: May 24, 2021
Related Concept Videos
Imaging Studies for Cardiovascular System V: CT
Radiological Investigation III: Pulmonary Angiogram and PET Scan
Pulmonary Angiogram
A Pulmonary Angiogram is an invasive procedure involving injecting a contrast medium through a catheter threaded into the pulmonary artery or the right side of the heart to visualize the pulmonary vasculature. Computed Tomography (CT) scans have mainly replaced this...
Radiological Investigation II: MRI and Ventilation Perfusion Scan
Magnetic Resonance Imaging (MRI) and Ventilation Perfusion Scans are two radiological investigations that offer detailed diagnostic images of the body, particularly lung structures.
MRI
MRI uses magnetic fields and radiofrequency signals to distinguish between normal and abnormal tissues. This technology provides a more detailed diagnostic image than CT scans, enabling it to characterize pulmonary nodules, stage bronchogenic carcinoma, and evaluate inflammatory activity in...