Related Experiment Video
Updated: Jun 25, 2026

Pressure Controlled Ventilation to Induce Acute Lung Injury in Mice
Published on: May 5, 2011
[Accidentally opened adjustable pressure-limiting valve. Failure of manual ventilation]
I Eggemann1, B W Böttiger, F Spöhr
1Klinik für Anästhesiologie und Operative Intensivmedizin, Universitätsklinik Köln, Köln.
Abstract:
In the case presented successful induction of anaesthesia was followed by failure to ventilate a patient during anaesthesia caused by a leakage in the anaesthesia circuit system during manual ventilation. The leakage problem could not be solved after checking the classical leakage sites (blocked tube, filter and tubes) and was due to a gas sensing line that caused a malfunction of the adjustable pressure-limiting valve.
Related Concept Videos
Ventilatory Modes
There are three ventilatory modes: full support, partial support, and spontaneous. These are described below.
Full Support Modes
Full support modes include controlled mechanical ventilation, continuous mandatory...
Mechanical Ventilation II: Invasive Ventilation
Negative-Pressure Ventilators
Negative-pressure ventilators create a vacuum around the chest or body to draw air into the lungs, simulating breathing. This method does not require an...
Mechanical Ventilation I: Indication and Settings
Mechanical Ventilation III: Noninvasive Ventilation
Noninvasive Positive-Pressure Ventilation (NIPPV)
Factors Affecting Pulmonary Ventilation
Alveolar Surface Tension
The alveolar fluid lines the luminal surface of the alveoli and exerts a force called surface tension. This force is caused by the polar water molecules in the liquid being more strongly attracted to each...
Pulmonary Ventilation: Inhalation
Boyle's law becomes particularly pertinent when examining respiratory...