Manufacture of a Multi-Purpose Low-Cost Animal Bench-Model for Teaching Tracheostomy
Published on: May 18, 2019
Automated emergency ventilation devices in a simulated unprotected airway
Holger Herff1, Christian A Schmittinger, Achim von Goedecke
1Department of Anesthesiology and Critical Care Medicine, Innsbruck Medical University, Innsbruck, Austria.
Background:
Automated ventilation devices are becoming more popular for emergency ventilation, but there is still not much experience concerning the optimal ventilation mode.
Methods:
In a bench model representing a non-intubated patient in respiratory and cardiac arrest, we compared a pressure-cycled with a time- and volume-cycled automated ventilation device in their completely automated modes. The main study endpoints were inspiratory time, respiratory rate, stomach inflation, and lung tidal volumes.
Results:
The pressure-cycled device inspired for 6.7 s in the respiratory arrest setting (respiratory rate 5.6/min), and never reached its closing pressure in the cardiac arrest setting (respiratory rate 1 breath/min). The time- and volume-cycled device inspired in both settings for 1.7 s (respiratory rate 13 breaths/min). In the respiratory arrest setting, mask leakage was 620 ± 20 mL for the pressure-cycled device vs. 290 ± 10 mL for the time- and volume-cycled device (p < 0.0001); lung tidal volume was 1080 ± 50 mL vs. 490 ± 20 mL, respectively (p < 0.0001); and there was no stomach inflation for either device. In the cardiac arrest setting, pressure-cycled device mask leakage was 5460 ± 60 mL vs. 240 ± 20 mL (p < 0.0001) for the time- and volume-cycled device (p < 0.0001); stomach inflation was 13,100 ± 100 mL vs. 90 ± 10 mL, respectively (p < 0.0001); and lung tidal volume 740 ± 60 mL vs. 420 ± 20 mL, respectively (p < 0.0001).
Conclusion:
In a simulated respiratory arrest setting, ventilation with an automated pressure-cycled ventilation device resulted in lower respiratory frequency and larger tidal volumes compared to a time- and volume-cycled device. In a simulated cardiac arrest setting, ventilation with an automated pressure-cycled ventilation device, but not a time- and volume-cycled device, resulted in continuous gastric insufflation.
Related Concept Videos
Cardiopulmonary Resuscitation II: ACLS Airway Management
Cardiopulmonary Resuscitation V: Advanced Airway Management Techniques
Ventilatory Modes
There are three ventilatory modes: full support, partial support, and spontaneous. These are described below.
Full Support Modes
Full support modes include controlled mechanical ventilation, continuous mandatory...
Mechanical Ventilation II: Invasive Ventilation
Negative-Pressure Ventilators
Negative-pressure ventilators create a vacuum around the chest or body to draw air into the lungs, simulating breathing. This method does not require an...
Mechanical Ventilation III: Noninvasive Ventilation
Noninvasive Positive-Pressure Ventilation (NIPPV)
Oxygen Delivering System II: Venturi Mask and Transtracheal Oxygen
Venturi Mask
The Venturi mask, named after the Venturi effect, is designed to deliver precise oxygen concentrations. It consists of a large tube with an oxygen inlet that narrows down, causing a pressure drop that pulls air in through adjustable side ports. The mask is a lightweight,...