Precision of flapless implant placement using real-time surgical navigation: a case series
Nicolas Elian1, Ziad N Jalbout, Anthony J Classi
1Division of Implant Dentistry, Department of Periodontology and Implant Dentistry, New York University College of Dentistry, New York, New York, USA.
Purpose:
To demonstrate the predictability of flapless surgery using navigation surgery.
Materials And Methods:
Computer-generated preoperative implant planning was compared to actual placement by CT (computerized tomography) scanning of patients before and after surgery. Once pre- and postoperative coordinates of virtual implants were obtained, linear distances and angles were calculated. Coronal and apical errors consisted of the shortest distance from the preoperative planning to the postoperative overlay.
Results:
Fourteen implants were placed in 6 patients who received CT scans before and after implant placement. Preoperative implant planning using software was compared to actual placement. The average discrepancy of the head of the implant was 0.89 mm +/- 0.53 SD (range, 0.32 to 1.96). The average discrepancy of the apex of the implant was 0.96 mm +/- 0.50 SD (range, 0.25 to 1.99). The average angle discrepancy and standard deviation were 3.78 degrees +/- 2.76 SD (range, 0.60 to 9.87).
Conclusion:
Optical computerized navigation is vulnerable to technological and technical errors. Yet, the present case series suggests that less than 1 mm of mean linear deviation and less than 4 degrees of angular deviation might be attainable.