Related Experiment Video
Updated: Jun 25, 2026

A New Murine Model of Endovascular Aortic Aneurysm Repair
Published on: July 7, 2013
Erosion of elephant trunk Dacron graft limb by thoracic endograft causing acute aneurysm expansion
John W Fehrenbacher1, Robert A McCready
1Department of Cardiovascular Surgery, Methodist Hospital, Clarian Health Partners, Indianapolis, Indiana, USA. jfehrenbacher@clarian.org
Abstract:
We recently treated a patient in whom a Gore TAG thoracic endograft (W.L. Gore and Assoc, Flagstaff, Arix) had been used to repair a descending thoracic aneurysm as the second stage of a hybrid procedure. This patient had previously undergone repair of ascending and aortic arch aneurysms, with an elephant trunk graft limb placed in the descending thoracic aorta for subsequent repair of the descending thoracic aneurysm. Eight months after placement of the thoracic endograft, the patient presented with an acutely expanding and symptomatic thoracic aneurysm. The patient was operated on urgently. The proximal portion of the endograft had eroded into the previously placed Dacron elephant trunk limb. The proximal portion of the endograft was removed and was replaced with a Dacron graft. The management of this patient forms the basis of this report.