Investigating the exposure class of a computed radiography system for optimisation of physical image quality for
C S Moore1, J R Saunderson, A W Beavis
1Radiation Physics Department, Hull and East Yorkshire Hospitals, Oncology Centre, Castle Hill Hospital, Cottingham, East Yorkshire. craig.moore@hey.nhs.uk
Abstract:
The purpose of this study was to investigate whether the exposure (speed) class (EC) of an Agfa computed radiography (CR) system could be used to optimise chest radiography. The frequency-dependent normalised noise-power spectra (NNPS(f)) were determined for a range of EC settings (25-1200) for a receptor dose of 4 microGy. Signal-to-noise ratios (SNRs) were measured in the lung, heart and diaphragm areas of a chest phantom with ECs of 400 and 600 at four tube voltages (60, 75, 90 and 125 kVp). As anatomical background can be a factor in detection of lung nodules, a tissue to rib ratio (TRR), which measures the ratio of pixel values in the nodule to that of rib, was measured in the lung region of the phantom to assess the suppression of the rib at ECs of 400 and 600. The NNPS(f) at ECs lower than 400 was relatively high. The NNPS(f) at EC 600 was found to be 7% lower when averaged over all frequencies than that at EC 400. The statistical significance of this difference was verified. The EC 800 and EC 1200 settings offered no extra advantages in terms of lowering frequency-dependent noise. The EC 600 setting offered improvements in SNR of between 10% and 18% in the lung, 11% and 16% in the heart, and 15% and 20% in the diaphragm compared with EC 400. Statistical analysis verified the significant difference. The EC 600 setting increased the TRR, thereby helping to suppress rib. This work indicates that an exposure class setting of 600 is the most appropriate for standard chest radiography, but clinical verification is required.
Related Concept Videos
Radiological Investigation I: X-ray and CT
X-ray Imaging
Computed Tomography
The technique was invented in the 1970s and is based on the principle that as X-rays pass through the body, they are absorbed or reflected at different levels. In the technique, a patient lies on a motorized platform while a computerized axial tomography (CAT) scanner rotates...
Imaging Studies for Cardiovascular System III: X-Ray
Definition and Purpose
An X-ray, or radiograph, is a non-invasive method that uses ionizing radiation to take images of internal structures. It is mainly used in cardiac imaging to examine the heart, lungs, and major blood vessels, aiming to identify abnormalities in the heart's size, shape, and position, such as heart failure, congenital defects, and vascular...
Positron Emission Tomography
One of the main requirements of a PET scan is a positron-emitting radioisotope, which is produced in a cyclotron and then attached to a substance used by the part of the body being...
Radiological Investigation III: Pulmonary Angiogram and PET Scan
Pulmonary Angiogram
A Pulmonary Angiogram is an invasive procedure involving injecting a contrast medium through a catheter threaded into the pulmonary artery or the right side of the heart to visualize the pulmonary vasculature. Computed Tomography (CT) scans have mainly replaced this...