Related Experiment Video
Updated: Jun 25, 2026

Visualization of Low-Level Gamma Radiation Sources Using a Low-Cost, High-Sensitivity, Omnidirectional Compton Camera
Published on: January 30, 2020
Large-area low-level gross alpha ZnS scintillation counting
1Radiometrisches Seminar, Universität, D-93040 Regensburg, Germany. henning.philipsborn@physik.uni-regensburg.de
Abstract:
An efficiency of (44+/-2)% is achieved for samples placed directly on a 5 in diameter ZnS scintillator plate, placed on top of a 5 in photomultiplier in a light-tight pot. Background count rate is as low as (9+/-3)h(-1), equivalent to (5+/-2)mBq on 95 cm(2) for a second scintillator face to face or for a high-purity copper disk. Results are displayed continuously and printed on-line. An automatic repetition of measurements permits a measurement of activity as a function of time. From such a function radionuclides such as radon and its decay products may be distinguished from other radionuclides by their half-life in a technically simpler way than by their energy.