Bioelectric Analyses of an Osseointegrated Intelligent Implant Design System for Amputees
Published on: July 15, 2009
A convenient approach for finite-element-analyses of orthopaedic implants in bone contact: modeling and experimental
Daniel Kluess1, Robert Souffrant, Wolfram Mittelmeier
1Department of Orthopaedics, University of Rostock, Rostock, Germany. daniel.kluess@med.uni-rostock.de
Abstract:
With regard to the growing potential of finite-element-analysis (FEA) in the field of orthopaedic biomechanics, we present an approach helping in the development of appropriate models of the implant-bone compound. The algorithm is based on computed-tomography data of the bone and accordant computer-aided-design (CAD) data of the implant and aims at predicting the bone strains and interface mechanics of the included parts. The developed algorithm was validated exemplary using an acetabular cup in combination with a left and a right fresh-frozen human hemipelvis. The strains under maximum loads during the gait cycle as well as the micromotion in the bone-implant interface were measured and compared to results from equivalent finite-element-analyses. Thereby, we found strong correlation between the calculated and measured principal strains with correlation coefficients of r(2)=0.94 (left side) and r(2)=0.86 (right side). A validation of micromotion was not possible due to limited accuracy of the motion tracking system.