A compact ammonia detector with sub-ppb accuracy using near-infrared photoacoustic spectroscopy and preconcentration sampling
Related Experiment Video
Updated: Jun 25, 2026

Fast and Accurate Exhaled Breath Ammonia Measurement
Published on: June 11, 2014
Andrea Pogány1, Arpád Mohácsi, Attila Varga
1Department of Optics and Quantum Electronics, University of Szeged, H-6701 P.O. Box 406 Szeged, Hungary. andrea@titan.physx.u-szeged.hu
Abstract:
This paper describes the first successful application of a preconcentration unitto increase the sensitivity of a photoacoustic ammonia concentration measuring instrument. A diode laser based near-infrared (1532 nm) photoacoustic ammonia monitoring instrument was combined with a tungsten (VI) oxide coated preconcentration unit to reach a sub-ppb detection limit with a compact, automatic measuring instrument. The system has no measurable cross-sensitivity to common atmospheric gases, most importantly to water vapor and carbon dioxide. The minimum detectable amount of ammonia is 2.9 ng, which means a minimum detectable concentration of 0.5 ppb with a 30-min measurementtime. Results of intercomparison measurements carried out both under laboratory and field conditions with reference to an electrochemical AMANDA instrument prove the applicability of the system in environmental ammonia concentration monitoring.
Related Concept Videos
Atomic Absorption Spectroscopy: Lab
Solutions containing organic solvents, such as low-molecular-mass alcohols, esters, or ketones, enhance absorbances by increasing nebulizer...
Inductively Coupled Plasma Atomic Emission Spectroscopy: Instrumentation
There are three main types of inductively coupled plasma atomic emission spectroscopy (ICP-AES) instruments: sequential, simultaneous multichannel, and Fourier transform instruments, with the latter being less commonly used.
Atomic Absorption Spectroscopy: Instrumentation
The atomizer used in AAS can be either a flame atomizer or an...
Gas Chromatography: Types of Detectors-II
Atomic Absorption Spectroscopy: Overview
When irradiated by EMR of a particular wavelength, these...
Atomic Emission Spectroscopy: Lab