Background:
Ambulatory oxygen is an important component of long-term oxygen therapy. Pulse-dose technology conserves oxygen and thus increases the operation time of a portable oxygen system.
Methods:
We tested 4 ambulatory oxygen systems (Helios, HomeFill, FreeStyle, and the compressed-oxygen cylinder system we regularly provide for long-term oxygen therapy at our Veterans Affairs hospital) with 39 subjects with stage-IV chronic obstructive pulmonary disease. Each subject performed one 6-min walk test with each oxygen system, and we measured blood oxygen saturation (via pulse oximetry [S(pO(2))]), heart rate, and modified Borg dyspnea score, and surveyed the subjects' preferences about the oxygen systems. We also studied whether the 2 systems that provide gas with a lower oxygen concentration (from a home concentrator or portable concentrator) showed any evidence of not providing adequate oxygenation.
Results:
With all 4 systems the mean pre-walk S(pO(2)) at the prescribed pulse-dose setting was 95-96%. The mean post-walk S(pO(2)) was 88-90% after each of the 4 walk tests. Between the 4 systems there were no statistically significant differences between the pre-walk-versus-post-walk S(pO(2)) ( = .42). With each system, the pre-walk-versus-post-walk S(pO(2)) difference was between -8% and -6%.
Conclusions:
Between these 4 ambulatory oxygen systems there were no significant differences in S(pO(2)), walk time, or walk distance, and there was no evidence of inadequate oxygenation with the 2 systems that provide a lower oxygen concentration.
More Related Videos
Related Concept Videos
Chronic Obstructive Pulmonary Disease-IV: Assessement and Diagnostic Studies
Medical History
Pulmonary Function Tests
Pulmonary Function Tests are crucial diagnostic tools for assessing respiratory function, particularly in patients with chronic respiratory disorders. They comprehensively evaluate lung volumes, ventilatory function, breathing mechanics, diffusion, and gas exchange. These tests help diagnose pulmonary diseases and play a significant role in monitoring disease progression, evaluating disability, and assessing response to therapy.
PFTs involve using a spirometer, a...
Treatment for Pulmonary Arterial Hypertension: Oxygen Therapy for Respiratory Failure
Oxygen therapy is vital in increasing and maintaining blood oxygen levels in PAH patients. As a result, it aids in reducing fatigue, improving...
Assessment of Respiration
Subjective Assessment: Nurses interview the patient to gather information directly during the subjective assessment. It includes questions about the individual's medical history, medications, and symptoms, focusing on past respiratory conditions like asthma or COPD,...
Special considerations while measuring oxygen saturation
Ensuring accuracy in vital sign recordings while prioritizing patient comfort and minimizing anxiety is important.
Respiratory Capacities
One key metric is the Inspiratory Capacity (IC), which represents the maximum amount of air that can be inhaled with full effort. IC is calculated by summing the tidal volume and inspiratory reserve volume, typically ranging from 2.4 to 3.6 liters.
The Functional Residual Capacity (FRC) represents the air in the...